logo

Standards Manage Your Business

We Manage Your Standards

ASTM

ASTM E1001 : 99a

Standard Practice for Detection and Evaluation of Discontinuities by the Immersed Pulse-Echo Ultrasonic Method Using Longitudinal Waves

Standard Details

1.1 This practice describes procedures for the ultrasonic examination of bulk materials or parts by transmitting pulsed, longitudinal waves through a liquid couplant into the material and observing the indications of reflected waves (Fig. 1). It covers only examinations in which one search unit is used as both transmitter and receiver (pulse-echo) and in which the part or material being examined is totally submerged in the couplant (immersion testing). This practice includes general requirements and procedures which may be used for detecting discontinuities and for making a relative or approximate evaluation of the size of discontinuities.

1.2 This practice complements Practice E214 by providing more detailed procedures for the selection and calibration of the inspection system and for evaluation of the indications obtained.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status : Historical
Standard Type: Main
Document No: ASTM E1001 : 99a
Document Year: 1999
Pages: 9
Edition: a
  • Section Volume:
  • 03.03 Volume 03.03 Nondestructive Testing (E94 – E2373)
  • ICS:
  • 19.100 Non-destructive testing

Life Cycle

Historical

Currently Viewing

Historical
ASTM E1001 : 99a
Knowledge Corner

Expand Your Knowledge and Unlock Your Learning Potential - Your One-Stop Source for Information!

© Copyright 2024 BSB Edge Private Limited.

Enquire now +