1.1 This practice describes the components of a direct current plasma (DCP) atomic emission spectrometer. This practice does not attempt to specify component tolerances or performance criteria. This practice does, however, attempt to identify critical factors affecting bias, precision, and sensitivity. A prospective user should consult with the vendor before placing an order to design a testing protocol for demonstrating that the instrument meets all anticipated needs.

1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific hazards statements are give in Section 9.

General Information

Status : Historical
Standard Type: Main
Document No: ASTM E1832 : 08(2012)
Document Year: 2008
Pages: 6
  • Section Volume:
  • 03.05 Volume 03.05 Analytical Chemistry for Metals, Ores, and Related Materials