logo

Standards Manage Your Business

We Manage Your Standards

ASTM

ASTM E1832 : 08(2017)

Standard Practice for Describing and Specifying a Direct Current Plasma Atomic Emission Spectrometer

Standard Details

1.1 This practice describes the components of a direct current plasma (DCP) atomic emission spectrometer. This practice does not attempt to specify component tolerances or performance criteria. This practice does, however, attempt to identify critical factors affecting bias, precision, and sensitivity. Before placing an order a prospective user should consult with the manufacturer to design a testing protocol for demonstrating that the instrument meets all anticipated needs.

1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific hazards statements are give in Section 9.

1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status : Historical
Standard Type: Main
Document No: ASTM E1832 : 08(2017)
Document Year: 2008
Pages: 6
  • Section Volume:
  • 03.05 Volume 03.05 Analytical Chemistry for Metals, Ores, and Related Materials

Life Cycle

Currently Viewing

Historical
ASTM E1832 : 08(2017)

Historical

Historical
Knowledge Corner

Expand Your Knowledge and Unlock Your Learning Potential - Your One-Stop Source for Information!

© Copyright 2024 BSB Edge Private Limited.

Enquire now +