ASTM E1582 : 14

Standard Practice for Calibration of Temperature Scale for Thermogravimetry

Standard Details

1.1 This practice describe the temperature calibration of thermogravimetric analyzers over the temperature range from 25 to 1500°C and is applicable to commercial and custom-built apparatus. This calibration may be accomplished by the use of either melting point StandardDetails or magnetic transition StandardDetails.

1.2 The mass change curve in thermogravimetry results from a number of influences, some of which are characteristic of the specimen holder assembly and atmosphere rather than the specimen. The variations from instrument to instrument occur in the point of measurement of the temperature, the nature of the material, its size and packing, the geometry and composition of the specimen container, the geometry and design of the furnace, and the accuracy and sensitivity of the temperature sensor and displaying scales. These all contribute to differences in measured temperatures, which may exceed 20°C. In addition, some sample holder assemblies will show variations of measured temperature with sample size or heating/cooling rate, or both. Since it is neither practical nor advisable to standardize sample holders or thermobalance geometries, instruments may be calibrated by measurement of the deviation of a melting or magnetic (Curie Point) transition temperature from the standard reference temperature. This deviation can be applied as a correction term to subsequent measurements.

1.3 This practice assumes that the indicated temperature of the instrument is linear over the range defined by a two-point calibration and that this linearity has been verified. These two calibration temperatures should be as close to the experimental measurements to be made as possible.

1.4 This practice describes two procedures for temperature calibration of thermogravimetric analyzers using any type balance. Procedure A uses melting point StandardDetails for calibration. Procedure B uses magnetic transition StandardDetails for calibration.

Note 1: Since all electronic data treatments are not equivalent, the user shall verify equivalency prior to use.

1.5 The data generated by these procedures can be used to correct the temperature scale of the instrument by either a positive or negative amount using either a two-point temperature calibration procedure or a multi-point temperature calibration with best line fit for the generated data.

Note 2: A single-point calibration may be used where this is the only procedure possible or practical. The use of a single-point procedure is not recommended.

1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.7 This practice is related to ISO 11358 but provides information and methods not found in ISO 11358.

1.8 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status : Historical
Standard Type: Main
Document No: ASTM E1582 : 14
Document Year: 2014
Pages: 7
  • Section Volume:
  • 14.01 Volume 14.01 Healthcare Informatics
  • ICS:
  • 17.200.20 Temperature-measuring instruments

Life Cycle

Historical

Currently Viewing

Historical
ASTM E1582 : 14